A power station for your CPAP: it's the humidifier that decides
Short answer: a CPAP machine itself sips power — around 11 watts with the humidifier off. A 500Wh station runs it for about 4 nights that way. But turn the heated humidifier up and add a heated tube, and the same machine can draw five times as much, cutting that to under 1 night. Size for the settings you actually sleep with.
The manufacturer's own numbers
ResMed publishes a battery guide for its AirSense 10 — the current each setup draws from a battery at each pressure. At a mid-range pressure of 10 cm H2O, converted to watts and multiplied over an 8-hour night, here is what one night of sleep actually costs:
| AirSense 10 setup (10 cm H2O) | Average draw | Energy per 8-hour night |
|---|---|---|
| No humidifier, standard tube | ~11 W | ~89 Wh |
| Humidifier on setting 4 | ~28 W | ~221 Wh |
| Humidifier on setting 8 | ~44 W | ~351 Wh |
| Humidifier on 8 + heated tube at 30°C | ~59 W | ~472 Wh |
Source: ResMed battery guide current figures for the AirSense 10 at 12V through ResMed's DC converter, checked November 2026. Other brands differ in detail, but every heated humidifier behaves the same way: heating water takes real power, and the blower barely takes any. Higher pressures draw somewhat more; lower pressures less.
So how many nights does a station last?
Using usable energy (rated Wh × 0.85, for inverter and reserve losses — the method explained in our runtime guide):
| Station size | Humidifier off (~89 Wh/night) | Humidifier mid (~221 Wh/night) | Humidifier max + heated tube (~472 Wh/night) |
|---|---|---|---|
| ~300Wh (usable ~255) | ~2.8 nights | ~1.1 nights | ~0.5 night |
| ~500Wh (usable ~425) | ~4.7 nights | ~1.9 nights | ~0.9 night |
| ~1,000Wh (usable ~870) | ~9.7 nights | ~3.9 nights | ~1.8 nights |
| ~2,000Wh (usable ~1,741) | ~19 nights | ~7.8 nights | ~3.7 nights |
Read the table by your worst realistic night, not your best. If you can sleep comfortably with the humidifier off or on a low setting during an outage, a modest station is genuinely multi-night backup. If you need full humidity and a heated tube, plan on roughly 500Wh of battery per night — and note ResMed itself builds a 50% safety margin into its own battery sizing recommendations, because mask leaks, cold rooms, and longer nights all push the draw up.
Three things that stretch a CPAP battery further
- Use the maker's DC converter instead of the wall plug. Running a CPAP from a station's AC outlet converts battery DC → AC → back to DC, wasting energy twice. ResMed's own guide recommends its DC/DC converter, and independent write-ups of ResMed's figures put the inverter penalty at roughly 15–30% more draw. On a multi-night outage, that's most of a free night.
- Turn the humidifier down or off for outage nights — or use passive humidification (water in the tank, heat off) if your clinician is fine with it. This is the single biggest lever there is: it can multiply your nights by three to five.
- Keep the station in the bedroom and the CPAP the only load. A station also running a fridge is doing two jobs; for CPAP-only backup, a smaller dedicated station by the bed is quieter, simpler, and always charged.
Which size should you buy for CPAP backup?
Humidifier off or low: a ~500Wh station is the sweet spot — several nights, light enough to travel with, and cheap insurance. Even ~300Wh covers a weekend.
Humidifier mid settings: look at ~1,000Wh for a comfortable 3–4 nights, or ~500Wh if one to two nights is all you need to bridge.
Full humidity + heated tube, non-negotiable: budget ~500Wh per night. A ~2,000Wh station (the class compared here) gives you about 3–4 nights and can share duty with the fridge — but if outages where you live run longer, solar recharging or a second battery matters more than brand choice.
A note on stakes: for many people a CPAP is medical equipment, not a convenience. This page helps you size a battery; it isn't medical advice, and it doesn't replace your clinician's guidance or your machine maker's instructions. If uninterrupted therapy is critical, size generously, keep the station charged, and test your setup on an ordinary night before you need it in a storm.